Section 3 - Electrical Connections and Setup Page 10
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
Section 3 - Electrical Connections and Setup
3.6 Paddle Wheel Flow Sensor connections and setup
Figure 3.5 Connections for Flow Sensor
Once connected, step through
the LCD menus to select the
sensor in the order shown. The
Sensor K factor (pulses per U.S.
Gallon) is usually printed on the
side of the sensor or on a label
attached to the sensor cable.
The drawing shows the connnections for a typical paddle wheel flow sensor. The cable to the
sensor should not exceed 2000’ (600 meters).
Dwg# N104-45
CAUTION:
Always disconnect
the loop power
before wiring or
unwiring a sensor